Please reset the UV meter to “0”J/cm2 before every time usage.
F358 is a UV broadband instrument, display with 4 UV bands:
UVA ( 320 – 400nm)
UVB ( 275 – 320nm)
UVC ( 200 – 275nm)
UVV ( 400- 700nm)
| F358 | |||
| Intensity Range | 0-30000mw/cm2 | Temp Range | -55℃∽+125℃ |
| Measurement range | 0-999J/cm2 | Test Accuracy | ±10% |
| Electric Power | 2 AAA alkaline dry batteries | Size | ∅118mm,H13mm |
| Weight | About 300g | Environmental Temp | 0-45℃ |
| Sampling speed | 2K Hz/s | Power data storage interval | 32 times/sec |
| Data storage time | minimum 4 minutes | Temperature data storage interval | 2 times/sec |
| Warranty | Free warranty for non-human damage within one year from the date of purchase, except for damage caused by objective factors and photoreceptor LCD filter consumables. | ||
| Calibration | Suggested to calibrate per year | ||
Key operation
- Power on/off
In the shutdown state, long press the “ON/OFF” button to start the operation.
In the power-on mode, and when the measurement mode is “MAX”, press and hold the “ON/OFF” key to power off.
When the measurement mode is “MAX”, without any operation, it will automatically shut down within 1 minute. Restart the computer and display the measurement results before the last shutdown.
In the “RT” state, the maximum waiting time is 50 minutes (Note: the maximum measurement recording time is 30 minutes).
- parameter settings
In the power-on state, long press “SELECT” to enter the calibration scale parameter modification setting mode:
In the setting mode, short press the “SELECT” key to select the modification position, short press the “ON/OFF” key to modify, after the modification, long press the “SELECT” key to save and return to the display page.